Block of detection BDZB-100B is intended for measuring the level of pollution of surfaces polluted by beta- active substances.

Output signal is a sequence of normalized, statistically distributed pulses, contains measurable data about density value(specific surface activity)of the flow of beta radiation where the block is located.

Area of the detector sensitive surface, cm2

80,0

Sensitivity*,·

(0,6±0,09) imp·s-1for1 min-1cm -2

Energy range of the registered beta-emission, MeV

From 0,3 to 3,5

Registration efficiency (2π)*, %

60,0

Own background, s-1

No more than 2

Type of the detector

Geiger-Muller Counter,Beta-5, 2 psc

Upper limit of measurement (with devices, produced by ZAO “Tetra”)— density of beta-emission flow*, min-1sm-2;

— specific service activity*, Bk·sm-2

2,0·106;

6,67·104 (doubled numerical value of the flow density, in s-1 • cm-2)

Upper limit of measurement (with other devices)— density of beta-emission flow*, min-1sm-2;

— specific service activity*, Bk·sm-2

2,0·104;

666,7

Overall dimensions, mm

210х150х50

Mass, kg

1,0

Connecting cable length, m

1,5 м

Note* – emission of 90Sr+90Y radionuclides

As the detection unit is equipped with a built-in object presence sensor, it greatly reduces the possibility of false operations while measuring the level of surface pollution.

ELECTRICAL CHARACTERISTICS

Power supply – DC voltage from 7,0 to 12,6 v. The consumed current is not more than 30 mA.

Designation of the joint contacts (plug RS -10) of the block is shown at the table.

contact

Signal designation

commentaries

1 + Uп 7,0 – 12,6 V
4 “Gamma”Output positive square-wave pulses with an amplitude of about  5 V, duration (1,5 ± 0,5) mks;
5 Sensor output Sensor of the object presence
9 Screen, connected to contact 10
10 General, connected to contact 9

 