Detection unit BDMG-100 is designed to measure the fields of gamma radiation in a wide range of dose rates.
Output signal – a sequence of normalized, statistically distributed pulses contains measuring information about the power of the equivalent of ambient dose H * (10) (PED) of gamma radiation.
Energy range of registered gamma rays MeV |
From 0,05 to 3,0 |
sensitivity*:
«sensitive» channel — «rough» channel — |
(4,4±0,44) с-1·for 1 mkZv·h-1 (5,0±2,0) с-1·for 1 mkZv·h-1 |
Measurement range** —
The upper limit of measured MED of gamma-radiation «sensitive» channel — «rough» channel — |
from 0.1 mkZv·h-1 to 10,0 Zv·h-1 2,0 mkZv·h-1 10,0 Zv·h-1 |
Own background, с-1 |
No more than 0,7 («sensitive» channel) |
Energy dependence of sensitivity *, % |
30,0 |
Detector type |
Geiger-Muller Counters, CBM-20 -2pcs; |
A control signal (switching of channels) |
Low level- voltage supply to the counters of both channels; |
Overall dimensions, mm |
Ø40×250 |
Mass, kg |
0,5 |
Connecting cable length, m |
20,0 (may be increased up to 200) |
Note: * — radiation of 137Cs radionuclide ** —optional-the upper limit of the measured gamma radiation MED of the “rough channel” can be increased up to 100.0 Zv • h-1 |
ELECTRICAL CHARACTERISTICS
Power supply – DC voltage from 7,0 to 12,6 v. The consumed current is not more than 30 mA. Variants of the detection unit: BDMG-100; BDMG -100-01; BDMG -100-02; BDMG -100-03.
Socket — plug RS-10.
Designation of contacts:
Contact |
Signal designation |
Commentaries |
|
1 | + Uп | 7,0 – 12,6 В | |
2 |
Control |
Set with an external device: |
|
4 |
Output of the “sensitive” channel |
Output of the “sensitive” channel: Square-wave pulses of about (1,5±0,5) mks Performance: |
|
5 |
Output of the “rough” channel. |
Output of the “rough” channel: Square-wave pulses of about (1,5±0,5) mks Performance: |
|
9 | ┴ | Screen , connected to contact 10 | |
10 | General | General, connected to contact 9 |
Variant of the detection unit — BDMG-100-07.
Socket — plug ОНЦ-БС-1-7/12.
Designation of contacts:
Contact |
Signal designation |
Commentary |
|
1 | + Uп | 7,0 – 12,6 В | |
2 | Control |
Set with an external device: |
|
4 | Output of the “sensitive” channel. | Output of the “sensitive” channel: positive square-wave pulses with an amplitude close to +Uп level, duration of (1,5±0,5) mks | |
5 | Output of the “rough” channel. | Output of the “rough” channel. positive square-wave pulses with an amplitude close to +Uп level, and duration of (1,5±0,5) mks | |
6 | ┴ | Screen | |
7 | General | General |