Block of detection BDZB-100 is intended for measuring the level of pollution of surfaces polluted by beta- active substances.
Output signal is a sequence of normalized, statistically distributed pulses, contains measurable data about the density value of the flow of beta radiation where the block is located.
Area of the detector sensitive surface, cm2 |
28,0 |
Sensitivity* |
(0,125±0,3) imp s-1 on 1 min-1∙sm-2 |
Energy range of the registered beta-emission, MeV |
From 0,3 to3,0 |
Registration efficiency (2π)*, % |
25,0 |
Type of the detector |
Scintillation plastic Ø68h2mm |
Upper limit of the measured density of beta-emission flow*, min-1·sm-2 |
1,0·105; |
Sensitivity to background gamma radiation |
No more than 19,2 imp s-1 on 1 mkZv/чh |
Unevenness of the sensitivity of the detector surface comparatively to the sensitivity in the center of the detector, % |
No more than 35 |
Overall dimensions, mm |
Ø90 х 220 |
Mass, kg |
1,0 |
Connecting cable length, m |
20,0 (may be increased up to 200 m) |
Note* – emission of 90Sr+90Y radionuclides |
ELECTRICAL CHARACTERISTICS
Power supply – DC voltage 12,6 v. The consumed current is not more than 30 mA.
Designation of the joint contacts (plug RS -10) of the block.
Contact |
Designation of the signal |
Commentaries |
|
1 |
+Uп | DC voltage +(12,6 ± 0,5) v | |
4 |
Output. | Output
BDZB-100: negative square-wave pulses with an amplitude of at least 4.5 V, duration (1,5 ± 0,5) s; BDZB-100-01: positive square-wave pulses with an amplitude of at least 4.5 V, duration (1,5 ± 0,5) s; BDZB-100-03: positive square-wave pulses with an amplitude of no less than 11,5 V, duration of (1,5 ± 0,5) s; |
|
9 | ┴ | Screen is connected to contact 10 | |
10 | General | General, connected to contact 9 |