Blocks of detection are intended  to measure the power of the ambient dose equivalent Å*(10) of gamma radiation and are used with free-running devices(BDMG-200UD; BDMG-200SD), and systems monitoring the radiological situation.

Output signal  corresponds to the standard interface:
• RS – 485 – detection unit BDMG-200 PD

• RS – 422 – detection unit BDMG-200 DD

• RS – 232 – detection unit BDMG-200 SD

• USB 1.1 – detection unit BDMG-200 UD

The contents and the format of the data are defined in the report of the information exchange DiBUS (www.doza.ru)

 

  1. Energy range of detected gamma radiation is from 9,6 to 480 fDzh (from 0,06 to 3,0 MeV)
    2. Measuring range of ambient dose equivalent H*(10) of gamma radiation is
    from 0.1 mkZv / h to 10 Zv / h:
    3. The measuring range is divided into two, automatically switching, sub-ranges:
    – Subrange “sensitive” channel – from 0.1 mkZv • h-1 to 2.0 mkZv • h-1;
    – Subrange “rough” channel – from 0.5 mkZv • h-1 to 10 Zv • h-1.
    4. Sensitivity *: “sensitive” channel – (4,4 ± 0,44) at 1 / Zv • h-1
    “Rough” channel – (5,0 ± 2,0) at 1 / mkZv • h-1.
    5. Limits of basic relative error of the block – ± 20%.
    6. Energy dependence * in the range from 9,6 to 200 fDzh (from 0,06 to 1,25 MeV) – no more than 30%
    7. Algorithm for processing measuring data – a method of “moving average”. Integration time is 40 seconds. Exposure – 1.
    8. Dimensions of the unit: diameter is 40 mm, length is 250 mm.
    9. Weight is not more than 0,5 kg.

* Radiation of radionuclide cesium-137 (106 fDzh).

ELECTRICAL CHARACTERISTICS

Power supply – is DC voltage from 7,0 to 12,6 V (detection unit BDMG-200UD – 5,0 V). Consumed current is not more than 10 mA.

Designation of the connector’s contacts (plug RS-10) block BDMG-200PD (RS-485).

Contact Signal designation Commentary
1 + Uп 7,0 – 12,6 V
2 f 1 Negative pulses from the output of the “rough” channel with an amplitude of about 5.0 V, and duration of 1.5 ms
3 f 2 Negative pulses from the output of the “sensitive” channel with an amplitude of about 5.0 V, and duration of 1.5 ms
4 TR+ Direct data
5 TR- Inverse data
9 Screen is connected to contact 10
10 General is connected to contact 9

Designation of the connector’s contacts (plug RS-10) block BDMG-200UD (USB 1.1).

Contact Signal designation Commentary
2 D- Inverse data
3 D+ Direct data
7 +Uп Supply voltage is +5V
9 Screen is connected to contact 10
10 General is connected to contact 9

Designation of the connector’s contacts (plug RS-10) block BDMG-200DD (RS-422).

Contact Signal designation Commentary
1 + Uп 7,0 – 12,6 V
4 Transfer P Transferring the direct signal
5 Transfer I Transferring the inversed signal
6 Receive P Receiving the direct signal
8 Receive I Receiving the inversed signal
9 Screen is connected to contact 10
10 General is connected to contact 9

Designation of connector’s contacts (plug RS-10) block BDMG-200SD (RS-232).

Contact Signal designation Commentary
1 + Uп 7,0 – 12,6 В
2
3 D Data
9 Screen is connected to contact 10
10 General is connected to contact 9

STRUCTURE OF DATA REGISTERS

Structure of DIBAS-data registers is shown in the tables 1.1 – 1.4

Table 1.1 – An array of data that can be written or read to / from BDMG-200

Index Designation Range Writing/Reading Type of the code
00 Measured data [0;∞] Reading/- 13
01 Efficiency of registration of the sensitive channel [0;∞] Reading/writing 13
02 Efficiency of registration of the rough channel [0;∞] Reading/writing 13
03 Threshold set point [0;∞] Reading/writing 13
00 Dead time of the sensitive channel [0;250] Reading/writing 1
01 Dead time of the rough channel [0;250] Reading/writing 1
02 Status [0;254] Reading/- 1
03 Network address of the device [0;254] -/writing 1
05 Switching the channels [0;2] Reading/writing 1

.